CALIBRATION CHAPTER 5
4. Enter the following code into the display using the Cursor Keys:
5111. Then press
( Left (), Right () to select digit ; Up (), Down () = increment/decrement value ;
Enter () to accept )
Calibrate tester
Using a laboratory millivolt meter, measure the voltage across pin 2 (ground) and 19 (Gage
Excitation) of the calibration fixture.
The display reads:
Example:
3.0211 volts
Program the measured voltage into the SURETEST Base Unit using the Cursor Keys. (Use Up (),
Down (), Left () and Right () Cursor Keys to match the SURETEST display to the millivolt-
meter). Then Press ENTER () to accept.
5. Next, the display reads:
Short pins 5 and 6 together on the calibration fixture. Then hit Enter () key.
6. Next, the display reads:
Remove the short between pins 5 and 6 of the calibration fixture.
Using a laboratory millivolt meter, measure the voltage across pins 5 and 6.
5-12