Snap-on 5000-3 User Manual Page 1

Browse online or download User Manual for Tools Snap-on 5000-3. Snap-On 5000-3 User Manual

  • Download
  • Add to my manuals
  • Print

Summary of Contents

Page 1 - SURETEST AND 600TL

SURETEST AND 600TL Model 5000-3 Torque Calibration System USER’S MANUAL

Page 2 - TABLE OF CONTENTS

FUNCTIONAL DESCRIPTION AND SPECIFICATIONS CHAPTER 2 CHAPTER 2 Functional Description and Specifications The CDI 5000-3 is CDI’s next generation Tor

Page 3

FUNCTIONAL DESCRIPTION AND SPECIFICATIONS CHAPTER 2 SURETEST BASE UNIT Front Panel Figure 2-2 SURETEST Front Panel A- Membrane Function

Page 4 - Using this Manual

FUNCTIONAL DESCRIPTION AND SPECIFICATIONS CHAPTER 2 Display A 240 x 64 Full-featured Graphic LCD is used to provide versatile and clear displays of s

Page 5 - CAUTION

FUNCTIONAL DESCRIPTION AND SPECIFICATIONS CHAPTER 2 The SURETEST returns to the default menu from the measurements mode in the event there is no selec

Page 6 - Safety Information

FUNCTIONAL DESCRIPTION AND SPECIFICATIONS CHAPTER 2 3. Select Language Selects either English, German (Deutsch), French (François), Spanish (Español)

Page 7

FUNCTIONAL DESCRIPTION AND SPECIFICATIONS CHAPTER 2 6. Edit Limits Sets up high torque limit preset. (-->) (F4) to go Down 1 line. Sets up low

Page 8 - Introduction

FUNCTIONAL DESCRIPTION AND SPECIFICATIONS CHAPTER 2 9. Edit Parameters Sets up serial communications protocol. Sets delay (in ms) per character u

Page 9

FUNCTIONAL DESCRIPTION AND SPECIFICATIONS CHAPTER 2 Sets up memory clear options in RECALL mode. Recalls memory data to display. Format (F3) calls S

Page 10 - CHAPTER 2

FUNCTIONAL DESCRIPTION AND SPECIFICATIONS CHAPTER 2 SURETEST BASE UNIT Rear Panel (Input/Output) Figure 2-3 SURETEST Rear Panel A - Solenoid

Page 11 - Front Panel

FUNCTIONAL DESCRIPTION AND SPECIFICATIONS CHAPTER 2 F- PRINTER PORT (DB-9P) PIN FUNCTION 2 Receive 3 Transmit 5 Ground G- OPTIONAL (DB-9P) PIN

Page 12 - Power Up Sequence

TABLE OF CONTENTS Safety Information I Introduction 1-1 Functional Description and Specifications 2-1 Functional Descriptions

Page 13 - Setup Mode

FUNCTIONAL DESCRIPTION AND SPECIFICATIONS CHAPTER 2 SURETEST Transducers SURETEST transducers provide industry standard square drives. They feature

Page 14

FUNCTIONAL DESCRIPTION AND SPECIFICATIONS CHAPTER 2 Specifications SURETEST TORQUE CALIBRATION SYSTEM Specifications System Accuracy ±0.25% of readin

Page 15

FUNCTIONAL DESCRIPTION AND SPECIFICATIONS CHAPTER 2 Power Requirements 100-230 VAC, 50-60 HZ, 50 W, automatic voltage selection. Fused at 3.15 A. E

Page 16

FUNCTIONAL DESCRIPTION AND SPECIFICATIONS CHAPTER 2 SURETEST Transducer/Display Resolution mkg cmkg ft. lb. in. lb. in. oz. Nm dNm cN

Page 17

SETUP AND PROGRAMMING CHAPTER 3 CHAPTER 3 SETUP AND PROGRAMMING CDI 5000-3 TORQUE CALIBRATION SYSTEM SETUP Figure 3-1: CDI 5000-3 Torque Tester/

Page 18 - Rear Panel (Input/Output)

SETUP AND PROGRAMMING CHAPTER 3 Setting Up the CDI 5000-3 Torque Calibration System 1. Bolt the loader to a sturdy location. 2. Install Pivot tub

Page 19 - F- PRINTER PORT (DB-9P)

SETUP AND PROGRAMMING CHAPTER 3 SURETEST Base Unit Controls Refer to the illustration below when performing the power-up and programming procedure

Page 20 - SURETEST Transducers

SETUP AND PROGRAMMING CHAPTER 3 4-in-1 Transducer Installation When a 4-in-1 transducer is installed, the following is displayed: The

Page 21 - Specifications

SETUP AND PROGRAMMING CHAPTER 3 Setup Programming Setting up Date and Time 1. Press (Setup). 2. Scroll Down (↓) to select Clock Adjus

Page 22

SETUP AND PROGRAMMING CHAPTER 3 3. Press Enter (↵). The display reads: 4. Press --> (F5) or <-- (F6) to select High or Low

Page 23 - SURETEST

SURETEST Base Unit Calibration 5-11 Gravitational Effects 5-14 Correction Factors on Test Weights 5-14 Gravitational Charts 5-1

Page 24 - SETUP AND PROGRAMMING

SETUP AND PROGRAMMING CHAPTER 3 To Program AUTO CLEAR, STORE, or SEND: 1. Press (Setup). 2. Scroll Down (↓) to select Auto settings

Page 25

SETUP AND PROGRAMMING CHAPTER 3 Clear Single Memory Location Use the procedure in this section to clear a single memory location. 1. Press

Page 26 - SURETEST Base Unit Controls

SETUP AND PROGRAMMING CHAPTER 3 1. Press Setup 2. Scroll Down (↓) to select Edit Parameters 3. Press Enter (↵) the display read: 4. Press --&

Page 27

TORQUE CALIBRATION SYSTEM SETUP CHAPTER 4 CHAPTER 4 CDI 5000-3 TORQUE CALIBRATION SYSTEM SETUP This chapter contains information on how to test,

Page 28 - Setup Programming

TORQUE CALIBRATION SYSTEM SETUP CHAPTER 4 Dial and Bending Beam Torque Wrenches and Screwdrivers Be sure the SURETEST transducer is capable of handl

Page 29 - Set-up AUTO Settings

TORQUE CALIBRATION SYSTEM SETUP CHAPTER 4 Adjustable or Preset “Click” Wrenches and Screwdrivers Be sure the SURETEST transducer is capable of handl

Page 30 - DATA LIST Memory

TORQUE CALIBRATION SYSTEM SETUP CHAPTER 4 Testing Power Tools The dynamic torque characteristic of a power tool and the static torque applied when u

Page 31

TORQUE CALIBRATION SYSTEM SETUP CHAPTER 4 Figure 4-1: Power Tool Test Setup A - Power Tool B - Joint Rate Simulator C - Transducer 7. Coupl

Page 32

TORQUE CALIBRATION SYSTEM SETUP CHAPTER 4 tool clutch slips. Be careful not to add torque by turning the tool by hand. 14. Release the torque an

Page 33 - CHAPTER 4

TORQUE CALIBRATION SYSTEM SETUP CHAPTER 4 Statistical Analysis The SURETEST Base Unit accumulates torque measurements in a data list. The list is c

Page 34 - WARNING

Using this Manual This manual contains instructions for use and setup of the SURETEST 5000-3 Torque Calibration System. A table of contents and a t

Page 35

TORQUE CALIBRATION SYSTEM SETUP CHAPTER 4 Printing Statistical Analysis 1. Press Data (F1) in select Transducer screen or (F2) in measurements mode

Page 36 - Non-Impact Tools

TORQUE CALIBRATION SYSTEM SETUP CHAPTER 4 ** R E S U L T ** Data : 0010 Max : 254.89 in-lb Min : 251.49 in-lb Range : 3.4039 in-lb Mean : 253

Page 37

TORQUE CALIBRATION SYSTEM SETUP CHAPTER 4 Displaying or Downloading Data List Torque measurements stored in the data list also include a date and t

Page 38

TORQUE CALIBRATION SYSTEM SETUP CHAPTER 4 How to Use Analog Output (Rear Panel BNC Connector) Analog Output on the SURETEST provides a real time vo

Page 39 - Statistical Analysis

CALIBRATION CHAPTER 5 CHAPTER 5 CALIBRATION The SURETEST Torque Calibration System is factory calibrated using precision torque test arms, ce

Page 40 - SPC PRINT-OUT

CALIBRATION CHAPTER 5 Torque Under 250 ft. lb. Mount the calibration block at the center of the stand with two 1/2-13 socket head cap screws w

Page 41

CALIBRATION CHAPTER 5 WARNING Always position the 40" arm over the front of the stand as shown. Never extend the test arm behind the sta

Page 42 - Displaying Data List

CALIBRATION CHAPTER 5 The 4-in-1 transducer may also be mounted to the calibration stand, shown in Figure 5-1, using the standoff fixture and

Page 43

CALIBRATION CHAPTER 5 Suggested Transducer Calibration Check Points, Bars, Hangers and Weights (Sheet 1 of 2) Part Bar Hanger

Page 44 - CALIBRATION

CALIBRATION CHAPTER 5 Suggested Transducer Calibration Check Points, Bars, Hangers and Weights (Sheet 2 of 2) Part Bar Hanger

Page 45 - Torque Over 250 ft. lb

DANGER Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury to the operator or to bystanders. W

Page 46

CALIBRATION CHAPTER 5 Torque Calibrations SURETEST New Transducer Calibration Improper calibration can result in torque measurement errors

Page 47 - Quick Check

CALIBRATION CHAPTER 5 6. Enter the following code into the display using the Cursor Keys: 5222 ( Left (←), Right (→) to select digit ; Up

Page 48

CALIBRATION CHAPTER 5 Press Unit (F2) key and select the desired torque unit. 10. Using the Up (↑) and Down (↓) Cursor keys, enter the Ful

Page 49

CALIBRATION CHAPTER 5 13. Press (Accept) to accept the calibration value (1st Cal point). Apply 20% Torque = 10 in. lb. 14. Press

Page 50

CALIBRATION CHAPTER 5 22. All readings show minus sign (-). Press F1 twice to return to measurement mode. SURETEST Base Unit Calibrat

Page 51 - Calibration CW:

CALIBRATION CHAPTER 5 4. Enter the following code into the display using the Cursor Keys: 5111. Then press (

Page 52

CALIBRATION CHAPTER 5 Example: 08.0622 mV 7. Program the measured voltage into the SURETEST Indicator as demonstrated in the example,

Page 53 - Calibration CCW:

CALIBRATION CHAPTER 5 Gravitational Effects Correction Factors on Test Weights Weights used in torque calibration are affected by their acce

Page 54 - Equipment

CALIBRATION CHAPTER 5 Gravitational Charts The charts in this section are taken from CRC Handbook of Chemistry and Physics, 58th Edition.

Page 55 - Calibrate tester

MANUAL LOADER CHAPTER 6 600TL - 600 ft. lb. Capacity Manual Loader The 600TL Manual Loader is designed as an integral component of the CDI 500

Page 56

SAFETY INFORMATION Safety Information Important Safety Instructions This manual contains important safety and operating instructions for CDI 5000-3

Page 57 - Gravitational Effects

MANUAL LOADER CHAPTER 6 Torque Tester/Calibrator Figure 6-1: Installation multiple transducers A- Quick release pin (Transducer) B- 4-in-1 T

Page 58 - Gravitational Charts

MANUAL LOADER CHAPTER 6 Drive Adaptor A drive adaptor couples the wrench to the transducer. Low torque 2000 series transducers use an internal-

Page 59 - Manual Loader

MANUAL LOADER CHAPTER 6 Figure 6-2 Set up for Dial indicating, Deflection Beam and Micro-Adjustable Torque Wrenches. 1- Transducer Mounting A

Page 60 - IMPORTANT

MANUAL LOADER CHAPTER 6 Types of Loader Testing Torque Driver Testing An optional torque screwdriver testing kit, is available for use in testing

Page 61

ACCESSORIES APPENDIX A Appendix A Accessories, included SURETEST 5000-3 Torque Calibration System Components Component Description 5000-ST SURE

Page 62

ACCESSORIES APPENDIX A Appendix A Accessories, optional Component Description 2000-4-02 1/4" ext. drive 5 – 50 in. oz. Transducer

Page 63 - Torque Driver Testing

ACCESSORIES APPENDIX A Component Description 2000-50-1 Cable, Indicator to PC 950-79 Foot Pedal Switch (Data Send foot swi

Page 64 - Accessories, included

ACCESSORIES APPENDIX A Calibration Accessories Calibration of the SURETEST Indicator and 2000 series transducers is recommended at least every 1 y

Page 65 - Accessories, optional

19220 San Jose Ave. City of Industry, CA 91748-1497 FORM 20-2100-CDI

Page 66 - Component Description

SAFETY INFORMATION so the torque and load values are indicated on the display. The safety relays only work when the power is on. Flying parti

Page 67 - Calibration Accessories

INTRODUCTION CHAPTER 1 CHAPTER 1 Introduction The SURETEST is a laboratory grade instrument that provides TORQUE measurements. Although desig

Page 68 - 19220 San Jose Ave

INTRODUCTION CHAPTER 1 The SURETEST operates directly from any AC power line between 100 VAC to 230 VAC, 50–60 Hz without the need for switch sele